Altair 4XR: Care and Use
Altair 4XR Multi-Gas Detector Guide
Charging the Altair 4XR
- Remove the Altair 4XR from the multi-unit charger.
- Green light indicates a fully charged device.
- Red light shows that charging is in progress.
- Briefly press the on/off button to switch on the instrument.
- The Altair 4XR will perform a self-test:
- All display segments will activate briefly.
- Acoustic, vibration alarms, and LED signals will be tested.
- The display will show:
- Manufacturer's name
- Product name
- Software version
- Bluetooth information
- Running sensor discovery
- Type of combustible gas being detected with measurement units
- Measurement units chosen for toxic gas
- Alarm set points for low and high alarms
- Short-term exposure limit
- Workplace exposure limit
- Calibration values
- Current time and date
- Date of the last calibration (if not deactivated) and due time in days for calibration (if activated).
- When "FAS" is displayed flashing, perform the fresh air setup by pressing the on/off button briefly.
Important: Do not activate the fresh air setup unless you are certain you are in an environment with fresh, uncontaminated air, otherwise inaccurate readings can occur.
Shutdown Procedure
- To shut down the instrument, press and hold the on/off button until "OFF" appears on the screen.
- Release the button, and the Altair 4XR will shut down.
Daily Bump Testing
- Insert the switched-on Altair 4XR into the Galaxy GX2 automatic test station to perform the daily bump test.
- The test normally takes approximately 40 seconds.
- A successful bump test will be indicated on the Galaxy GX2 display.
- Remove the Altair 4XR from the test station.
- The instrument's display will show a check mark for the next 24 hours after successful bump testing.
- The Altair 4XR is now ready to use.
Note: The Galaxy GX2 will inform you if the bump test and subsequent calibration are not successful. Take out the Altair 4XR if this occurs. The multi-gas detector will not be ready to use and will require an inspection.
Low Battery Warning
- When the Altair 4XR enters low battery warning, the battery life indicator flashes, alarms sound, and alarm LEDs flash.
- The display shows "LOW BAT".
- Leave the area immediately as the end of battery life is approaching.
Bluetooth Communication
- Bluetooth communication must be enabled to operate with the mobile device and installed Altair Connect app.
- The instrument will automatically enter discovery mode for five minutes after the device startup sequence.
- To manually enter discovery mode:
- Page down through the menu pages in normal operation mode using the arrow button until the discovery mode page is displayed.
- Press the up arrow button until the display shows "DISCOVERY YES".
- Press the on/off button to enter discovery mode.
- The blue LED will blink rapidly, indicating that the device is in discovery mode.
Connecting to Altair Connect App
- Start the Altair Connect app and press "Search for Instruments" on the device.
- Tap the "Pair" button for the displayed instrument.
- Both the instrument and device will display a matching unique six-digit security code.
- Confirm the pairing request on both the device and the Altair 4XR by pressing the arrow down button.
- The Altair 4XR also has an integrated RFID chip for faster Bluetooth pairing.
- If you have successfully established a Bluetooth connection, the Altair Connect app will display the instrument readings, and the blue LED on the instrument will flash slowly.
Safety IO Grid Live Monitor Subscribers: You will see the gas readings and location details of the connected user.
Disconnecting from the App
- To disconnect the instrument from the app, tap the three dots in the upper right corner of the device and tap "Disconnect".